HP CODEMASTER XL:
The Defibrillator Chosen Most By Medical Professionals

Overview: The HP CodeMaster XL comes with
ECG, recorder, all cables, manual and new batteries,
with a 90-day warranty.
This versatile, high-quality unit charges from 0
to 360 joules in under 5 seconds. Plus it's packed
with standard features you normally won't find in
a defibrillator/monitor at this price.
The CodeMaster XL provides true 1-2-3 operation,
built-in pediatric paddles under slide-off adult paddles,
preset heart rate alarms, a high-capacity battery,
built-in diagnostics, and auto gain--which automatically
sizes ECGs on its large, 5-inch display.
Not
only does the CodeMaster XL deliver all the speed
and performance you need in general care, it offers
you a variety of advanced options: built-in Pulse
Oximetry (SpO2), Shock Advisory with Advisory Event
Summary, a paddle-to-patient contact indicator, and
hands-free defibrillation.
Standard Features
- Charges from 0 to 360 joules in less than five
seconds.
- True 1-2-3 operation, standardized with previous
HP defibrillators.
- Advisory Event Summary saves approximately 200
events and up to 50 ECG strips, with data retained
after unit is turned off (with Shock Advisory option).
- Full range of internal paddles available.
- Built-in pediatric paddles available under slide-off
adult electrodes.
- Optional hands-free defibrillation available.
- LEADS OFF message appears on monitor if a lead
becomes disconnected.
Advanced Features
- Optional shock advisory capability allows use
as an advisory or semi-automatic defibrillation.
- Optional Pulse Oximeter(SpO2).
- Optional high-visibility yellow case for easy
recognition.
- Paddle-to-patient contact indicator reduces guesswork
(Optional on CodeMaster XL).
- Adjustable heart rate alarms (Preset on CodeMaster
XL).
- Higher capacity battery powers 50 full-energy
discharges. Also powers 2.5 hours ECG monitoring.
- Optional built-in noninvasive pacing (CodeMaster
XL+).
- Event Summary documents pre- and post-shock data
(CodeMaster XL+).
Detailed Specifications
| Dimensions |
7.9"H x 11.8"W x 15.6"L/20
x 30 x 39.7 cm |
| Weight |
25 lbs / 10.9 kg |
| Battery |
Type: Rechargeable, sealed
lead-acid battery. 4 Ah, 12 V nominal.
Charge Time: 2 hours to 90% of full capacity.
18 hours to 100% capacity. Repeated chaging to
less than 100% will reduce the useful life of
the battery.
Capacity: 2.5 hours monitoring or 50 full-energy
discharges or 1 hour monitoring and recording.
AC input: 100 to 230 VAC 15% 50/60 Hz.
Battery Indicators: illuminated LED indicates
battery is charging. Low Battery message appears
on the monitor. |
| Defibrillator |
Waveform: damped sinusoidal.
Output energy (delivered): 2, 3, 5, 7,
10, 20, 30, 50, 70, 100, 150, 200, 300 and 360
joules.
Charge Control: push-button on apex paddle and
on front panel
Charge Time: Less than 5 seconds to 360
joules with a functional battery installed.
Charge-Done Indicator: Display indicates
available energy, charge-done light on apex paddle,
charge done tone.
Synchronizer: SYNC message appears on the
monitor and is annoted periodically on the recorder
while in synchronous mode. An audible beep sounds
with each detected R-wave, while a marker on the
monitor and sync mark on the recorder strip indicate
the discharge point.
Paddles: standard paddels are anterior/anterior.
Adult electrodes (83 cm sq) slide off to expose
pediatric electrodes (21 cm sq). Paddle cord is
10 feet (3m). Full range of internal paddles available.
|
| Monitor |
Input: ECG Activity may be
viewed through the paddles or patient cable.
Lead Fault: LEADS OFF message and dashed
baseline appear on the monitor if a lead becomes
disconnected.
Common Mode Rejection: Greater than 100
dB measured as per AAMI standards for cardiac
monitors
Heart Rate Display: Digital read-out on
monitor from 20 to 280 bpm. |
| Recorder |
Annotations: time, date,
heart rate, event markers, ECG mode, and selected
energy.
Recorder Mode: Automatically documents
events and ECG during defibrillation episodes.
The recorder can be configured to run in either
real time or with a 6-second delay
Frequency Response: 0.5 to 40 Hz monitor
bandwidth. 0.005 to 150 Hz diagnostic bandwidth
is selectable. |
|